Alcohol (Blood Ethanol): Survey Deficiencies and CLIA Citations
Most Blood Alcohol citations come from five places: quality control (42 CFR 493.1256), calibration verification (42 CFR 493.1255), proficiency testing or accuracy verification, method verification (42 CFR 493.1253) and test reporting, including critical values (42 CFR 493.1291). Keep the records for each area current and you remove most of the risk.
Where Blood Alcohol citations come from
Survey findings about Blood Alcohol almost always trace to a short list of 42 CFR 493 standards. This table shows each one and the records a surveyor will usually ask to see.
| Area | Regulation | Records surveyors ask for |
|---|---|---|
| QC | 42 CFR 493.1256 | Daily Blood Alcohol QC records, Levey-Jennings charts, corrective action log, control lot ranges |
| Calibration verification | 42 CFR 493.1255 | Six-month Blood Alcohol calibration verification and reportable range data with director review |
| Proficiency testing | 42 CFR 493.801, Subpart H | Blood Alcohol PT enrollment, attestation statements, graded reports, investigation of every miss |
| Method verification | 42 CFR 493.1253 | Blood Alcohol verification study: accuracy, precision, reportable range, reference interval check, director sign-off |
| Test report and critical values | 42 CFR 493.1291 | Critical value list, call log with read-back, report format showing Blood Alcohol reference intervals |
| Personnel | Subpart M | Training and competency records for everyone who runs Blood Alcohol |
PT status for Blood Alcohol
Alcohol (Blood Ethanol) is a regulated analyte (toxicology, 42 CFR 493.937). CLIA PT acceptance limit: target Value ±20%.
Specimen problems that turn into Blood Alcohol findings
Preanalytical errors become citations when the lab has no written rule for them, or does not follow its own rule.
- Do not clean the site with alcohol swabs; use a non-alcohol antiseptic.
- Keep tubes capped; ethanol evaporates from open containers.
- Medical serum/plasma results are not the same as whole blood forensic results; state specimen type.
If Blood Alcohol was cited on your survey
- Identify the exact standard cited and the records the surveyor reviewed.
- Look back at patient results that could have been affected and decide whether corrected reports are needed.
- Fix the system, not just the instance, and assign monitoring.
- Write the plan of correction. See the CLIA plan of correction guide.
Related tests
- Acetaminophen (Serum) (full guide)
- Salicylate (full guide)
- Urine Drug Screen (Drugs of Abuse) (full guide)
Full guide: Alcohol (Blood Ethanol) testing in the clinical lab: CLIA, PT and QC requirements on soundlabconsult.com, part of the Laboratory Test Compliance Library.
Cited on a CLIA survey?
Sound Lab Consultants helps laboratories read their citations, find root causes and write plans of correction.
Related guides on soundlabconsult.com:
Frequently asked questions
What are the most common Blood Alcohol survey deficiencies?
Missing or undocumented QC, missed calibration verification, PT or accuracy verification gaps, no method verification study, and undocumented critical value calls for Blood Alcohol. These are general categories, not statistics.
Which regulation covers Blood Alcohol quality control?
42 CFR 493.1256 sets the control procedure requirements.
Is Blood Alcohol on the CLIA PT analyte list?
Alcohol (Blood Ethanol) is a regulated analyte (toxicology, 42 CFR 493.937). CLIA PT acceptance limit: target Value ±20%.
Sources
- 42 CFR 493.937, Toxicology
- 42 CFR 493.1256, Control procedures
- 42 CFR 493.1255, Calibration and calibration verification
- 42 CFR 493.1253, Establishment and verification of performance specifications
- 42 CFR 493.1291, Test report
- 42 CFR 493.1236, Evaluation of proficiency testing performance
- CMS CLIA Year 2024 Top 10 Standard- and Condition-Level Deficiencies
- CMS State Operations Manual, Appendix C (Laboratories)