ALT (Alanine Aminotransferase): Survey Deficiencies and CLIA Citations
Most ALT citations come from five places: quality control (42 CFR 493.1256), calibration verification (42 CFR 493.1255), proficiency testing or accuracy verification, method verification (42 CFR 493.1253) and test reporting, including critical values (42 CFR 493.1291). Keep the records for each area current and you remove most of the risk.
Where ALT citations come from
Survey findings about ALT almost always trace to a short list of 42 CFR 493 standards. This table shows each one and the records a surveyor will usually ask to see.
| Area | Regulation | Records surveyors ask for |
|---|---|---|
| QC | 42 CFR 493.1256 | Daily ALT QC records, Levey-Jennings charts, corrective action log, control lot ranges |
| Calibration verification | 42 CFR 493.1255 | Six-month ALT calibration verification and reportable range data with director review |
| Proficiency testing | 42 CFR 493.801, Subpart H | ALT PT enrollment, attestation statements, graded reports, investigation of every miss |
| Method verification | 42 CFR 493.1253 | ALT verification study: accuracy, precision, reportable range, reference interval check, director sign-off |
| Test report and critical values | 42 CFR 493.1291 | Critical value list, call log with read-back, report format showing ALT reference intervals |
| Personnel | Subpart M | Training and competency records for everyone who runs ALT |
PT status for ALT
ALT (Alanine Aminotransferase) is a regulated analyte (routine chemistry, 42 CFR 493.931). CLIA PT acceptance limit: target value ±15% or ±6 U/L (greater).
Specimen problems that turn into ALT findings
Preanalytical errors become citations when the lab has no written rule for them, or does not follow its own rule.
- Hemolysis (in vitro) during collection or transport can falsely raise results; set a hemolysis index rejection or comment rule.
- Delayed separation and storage temperature affect stability; follow the manufacturer's stability claims.
- Pyridoxal phosphate (P5P)-activated vs. non-activated methods can disagree; keep the same method family when comparing results.
If ALT was cited on your survey
- Identify the exact standard cited and the records the surveyor reviewed.
- Look back at patient results that could have been affected and decide whether corrected reports are needed.
- Fix the system, not just the instance, and assign monitoring.
- Write the plan of correction. See the CLIA plan of correction guide.
Related tests
- AST (Aspartate Aminotransferase) (full guide)
- Alkaline Phosphatase (ALP) (full guide)
- Bilirubin, Total (full guide)
- GGT (Gamma-Glutamyl Transferase) (full guide)
- Albumin (full guide)
Full guide: ALT (Alanine Aminotransferase) testing in the clinical lab: CLIA, PT and QC requirements on soundlabconsult.com, part of the Laboratory Test Compliance Library.
Cited on a CLIA survey?
Sound Lab Consultants helps laboratories read their citations, find root causes and write plans of correction.
Related guides on soundlabconsult.com:
Frequently asked questions
What are the most common ALT survey deficiencies?
Missing or undocumented QC, missed calibration verification, PT or accuracy verification gaps, no method verification study, and undocumented critical value calls for ALT. These are general categories, not statistics.
Which regulation covers ALT quality control?
42 CFR 493.1256 sets the control procedure requirements.
Is ALT on the CLIA PT analyte list?
ALT (Alanine Aminotransferase) is a regulated analyte (routine chemistry, 42 CFR 493.931). CLIA PT acceptance limit: target value ±15% or ±6 U/L (greater).
Sources
- 42 CFR 493.931, Routine chemistry
- 42 CFR 493.1256, Control procedures
- 42 CFR 493.1255, Calibration and calibration verification
- 42 CFR 493.1253, Establishment and verification of performance specifications
- 42 CFR 493.1291, Test report
- 42 CFR 493.1236, Evaluation of proficiency testing performance
- CMS CLIA Year 2024 Top 10 Standard- and Condition-Level Deficiencies
- CMS State Operations Manual, Appendix C (Laboratories)