CEA (Carcinoembryonic Antigen): Survey Deficiencies and CLIA Citations

Most CEA citations come from five places: quality control (42 CFR 493.1256), calibration verification (42 CFR 493.1255), proficiency testing or accuracy verification, method verification (42 CFR 493.1253) and test reporting, including critical values (42 CFR 493.1291). Keep the records for each area current and you remove most of the risk.

Last reviewed Nov 24, 2026 · By the Sound Lab Consultants regulatory team · Sources listed below

Where CEA citations come from

Survey findings about CEA almost always trace to a short list of 42 CFR 493 standards. This table shows each one and the records a surveyor will usually ask to see.

AreaRegulationRecords surveyors ask for
QC42 CFR 493.1256Daily CEA QC records, Levey-Jennings charts, corrective action log, control lot ranges
Calibration verification42 CFR 493.1255Six-month CEA calibration verification and reportable range data with director review
Proficiency testing42 CFR 493.801, Subpart HCEA PT enrollment, attestation statements, graded reports, investigation of every miss
Method verification42 CFR 493.1253CEA verification study: accuracy, precision, reportable range, reference interval check, director sign-off
Test report and critical values42 CFR 493.1291Critical value list, call log with read-back, report format showing CEA reference intervals
PersonnelSubpart MTraining and competency records for everyone who runs CEA

PT status for CEA

CEA (Carcinoembryonic Antigen) is a regulated analyte (endocrinology, 42 CFR 493.933). CLIA PT acceptance limit: target value ±15% or ±1 ng/mL (greater).

Specimen problems that turn into CEA findings

Preanalytical errors become citations when the lab has no written rule for them, or does not follow its own rule.

If CEA was cited on your survey

  1. Identify the exact standard cited and the records the surveyor reviewed.
  2. Look back at patient results that could have been affected and decide whether corrected reports are needed.
  3. Fix the system, not just the instance, and assign monitoring.
  4. Write the plan of correction. See the CLIA plan of correction guide.

Related tests

Full guide: CEA (Carcinoembryonic Antigen) testing in the clinical lab: CLIA, PT and QC requirements on soundlabconsult.com, part of the Laboratory Test Compliance Library.

For laboratory compliance and operations only, not medical advice or patient care guidance.

Cited on a CLIA survey?

Sound Lab Consultants helps laboratories read their citations, find root causes and write plans of correction.

Book a free consultation

Related guides on soundlabconsult.com:

Frequently asked questions

What are the most common CEA survey deficiencies?

Missing or undocumented QC, missed calibration verification, PT or accuracy verification gaps, no method verification study, and undocumented critical value calls for CEA. These are general categories, not statistics.

Which regulation covers CEA quality control?

42 CFR 493.1256 sets the control procedure requirements.

Is CEA on the CLIA PT analyte list?

CEA (Carcinoembryonic Antigen) is a regulated analyte (endocrinology, 42 CFR 493.933). CLIA PT acceptance limit: target value ±15% or ±1 ng/mL (greater).