SARS-CoV-2 (COVID-19) Testing: Survey Deficiencies and CLIA Citations
Most COVID-19 citations come from five places: quality control (42 CFR 493.1256), calibration verification (42 CFR 493.1255), proficiency testing or accuracy verification, method verification (42 CFR 493.1253) and test reporting, including critical values (42 CFR 493.1291). Keep the records for each area current and you remove most of the risk.
Where COVID-19 citations come from
Survey findings about COVID-19 almost always trace to a short list of 42 CFR 493 standards. This table shows each one and the records a surveyor will usually ask to see.
| Area | Regulation | Records surveyors ask for |
|---|---|---|
| QC | 42 CFR 493.1256 | Daily COVID-19 QC records, Levey-Jennings charts, corrective action log, control lot ranges |
| Proficiency testing | 42 CFR 493.801, Subpart H | COVID-19 PT enrollment, attestation statements, graded reports, investigation of every miss |
| Method verification | 42 CFR 493.1253 | COVID-19 verification study: accuracy, precision, reportable range, reference interval check, director sign-off |
| Test report and critical values | 42 CFR 493.1291 | Critical value list, call log with read-back, report format showing COVID-19 reference intervals |
| Personnel | Subpart M | Training and competency records for everyone who runs COVID-19 |
PT status for COVID-19
No analyte-level limit; nonwaived labs enroll in virology PT for the services they perform (42 CFR 493.919).
Specimen problems that turn into COVID-19 findings
Preanalytical errors become citations when the lab has no written rule for them, or does not follow its own rule.
- Specimen collection technique and transport media drive false negatives; train collectors.
- Antigen tests have lower sensitivity than NAAT; follow the instructions for confirmatory testing.
- Public health reporting requirements may apply.
If COVID-19 was cited on your survey
- Identify the exact standard cited and the records the surveyor reviewed.
- Look back at patient results that could have been affected and decide whether corrected reports are needed.
- Fix the system, not just the instance, and assign monitoring.
- Write the plan of correction. See the CLIA plan of correction guide.
Related tests
- Influenza A/B Testing (full guide)
- RSV (Respiratory Syncytial Virus) Testing (full guide)
- Group A Strep Testing
Full guide: SARS-CoV-2 (COVID-19) Testing testing in the clinical lab: CLIA, PT and QC requirements on soundlabconsult.com, part of the Laboratory Test Compliance Library.
Cited on a CLIA survey?
Sound Lab Consultants helps laboratories read their citations, find root causes and write plans of correction.
Related guides on soundlabconsult.com:
Frequently asked questions
What are the most common COVID-19 survey deficiencies?
Missing or undocumented QC, missed calibration verification, PT or accuracy verification gaps, no method verification study, and undocumented critical value calls for COVID-19. These are general categories, not statistics.
Which regulation covers COVID-19 quality control?
42 CFR 493.1256 sets the control procedure requirements.
Is COVID-19 on the CLIA PT analyte list?
No analyte-level limit; nonwaived labs enroll in virology PT for the services they perform (42 CFR 493.919).
Sources
- 42 CFR 493.919, Virology
- 42 CFR 493.1256, Control procedures
- 42 CFR 493.1255, Calibration and calibration verification
- 42 CFR 493.1253, Establishment and verification of performance specifications
- 42 CFR 493.1291, Test report
- 42 CFR 493.1236, Evaluation of proficiency testing performance
- CMS CLIA Year 2024 Top 10 Standard- and Condition-Level Deficiencies
- CMS State Operations Manual, Appendix C (Laboratories)