Accrediting organization citations and condition-level findings
An accrediting organization cites a lab against its own standards, which CMS requires to be at least as strict as CLIA condition-level requirements. Most findings stay inside the AO's correction process, but some reach CMS: the AO must report immediate jeopardy within 10 days and adverse actions within 30 days (42 CFR 493.555), and condition-level noncompliance found on a CMS validation inspection puts the lab under CMS enforcement (42 CFR 493.569).

How accreditation standards line up with CLIA conditions
To be approved, an accreditation organization must give CMS a crosswalk of its standards to CLIA and show they are equal to, or more stringent than, the CLIA condition-level requirements (42 CFR 493.553, 42 CFR 493.555). So an AO checklist citation usually points back to a CLIA subpart: proficiency testing (Subpart H), facility administration (Subpart J), quality systems (Subpart K) or personnel (Subpart M).
Which accreditation findings reach CMS
| Situation | What happens | Rule |
|---|---|---|
| Deficiency posing immediate jeopardy or a public hazard | AO notifies CMS within 10 days | 42 CFR 493.555(c)(2) |
| Accreditation suspended, withdrawn, revoked or limited, or lab sanctioned | AO notifies CMS within 30 days | 42 CFR 493.555(c)(1) |
| Unsuccessful proficiency testing | Results released to CMS; AO reports its action within 30 days | 42 CFR 493.551(b)(5)-(6) |
| Condition-level noncompliance found on a CMS validation inspection | Full CMS review; principal and alternative sanctions possible | 42 CFR 493.569 |
| Lab refuses a validation inspection | Full CMS review; certificate of accreditation may be suspended, revoked or limited | 42 CFR 493.567 |
Condition level vs standard level
Under 42 CFR 493.2, a condition-level deficiency is noncompliance with one or more condition-level requirements. Standard-level findings are corrected through a plan of correction; condition-level findings can bring sanctions. Accredited labs face the same distinction when CMS validates their accreditation.
Are accreditation citations public?
CMS may disclose AO inspection results only when related to an enforcement action, but may disclose the results of all its validation inspections (42 CFR 493.571). CLIA 2567s from state and CMS surveys are searchable in QCOR.
The CMS-approved CLIA accreditors
CMS currently approves seven accreditation organizations under CLIA: AABB, A2LA, ACHC, ASHI, CAP, COLA and The Joint Commission (CMS list v. 8.8.2025). A2LA, ACHC, CAP, COLA and The Joint Commission are approved for all 24 specialties and subspecialties on CMS's September 2026 matrix; AABB for 14 and ASHI for 4.
Cited by your accreditor?
See the full list of CMS-approved accrediting organizations by program and the CLIA accreditation organizations and their approved specialties on soundlabconsult.com, with CMS source dates for every list. Book a free consultation with Sound Lab Consultants, an independent consulting firm not affiliated with CMS or any accrediting organization.
Cited on a CLIA survey?
Sound Lab Consultants helps laboratories read their citations, find root causes and write plans of correction.
Related guides on soundlabconsult.com:
Frequently asked questions
Do CAP or COLA citations go to CMS?
Not routinely. The accrediting organization must notify CMS of immediate jeopardy deficiencies within 10 days and of adverse actions such as suspension or limitation within 30 days (42 CFR 493.555).
Can an accreditation citation lead to CLIA sanctions?
Yes, if CMS finds condition-level noncompliance on a validation inspection (42 CFR 493.569), or after an adverse action reported by the accreditor.
What is a condition-level deficiency?
Noncompliance with one or more condition-level requirements of 42 CFR 493, as defined in 42 CFR 493.2.
Which accreditors does CMS approve under CLIA?
AABB, A2LA, ACHC, ASHI, CAP, COLA and The Joint Commission (CMS list v. 8.8.2025).
Are accreditation inspection results public?
CMS may disclose them only if related to an enforcement action (42 CFR 493.571); CMS validation inspection results may be disclosed.
Sources
- 42 CFR 493 Subpart E, Accreditation by a Private, Nonprofit Accreditation Organization
- CMS, List of Approved Accreditation Organizations under CLIA (v. 8.8.2025)
- CMS, Specialty/Subspecialty Information for Approved Accreditation Organizations under CLIA (September 2026)
- 42 CFR Part 493 (eCFR)
- CMS QCOR (Quality, Certification and Oversight Reports)
- CMS, Accrediting Organizations (AOs) page (last modified 08/17/2026)