CLIA citation FAQ
Short answers, each tied to the regulation. Follow the links for the full guide.
Last reviewed Nov 24, 2026 · By the Sound Lab Consultants regulatory team
Book a free consultationGuides
Frequently asked questions
What is a D-tag in a CLIA survey?
A tag number beginning with D that identifies the CLIA requirement a surveyor cited. CMS lists the tags, regulation text and interpretive guidelines in SOM Appendix C.
What is the difference between a condition-level and standard-level deficiency?
A condition-level deficiency is noncompliance with one or more condition level requirements (42 CFR 493.2). Standard-level findings require an acceptable plan of correction (493.1816); condition-level findings can lead to sanctions (493.1806).
Can a CLIA citation lead to losing a certificate?
Suspension, limitation and revocation of the CLIA certificate are the principal sanctions under 42 CFR 493.1806. They are tied to condition-level noncompliance, not to every standard-level finding.
Are CLIA survey citations public?
CMS publishes CLIA survey information through its QCOR reporting site.
Where can I look up a CLIA D-tag?
In CMS State Operations Manual Appendix C, which lists each D-tag with the 42 CFR 493 regulation text and interpretive guidelines.
What is D2016?
In Appendix C, D2016 is the tag for 42 CFR 493.803, Condition: Successful participation in proficiency testing.
What is D5400?
D5400 is the Appendix C tag for 42 CFR 493.1250, Condition: Analytic systems.
What is a CMS-2567?
The Statement of Deficiencies and Plan of Correction, the form on which CLIA surveyors record deficiencies and the lab records its plan of correction.
How long do I have to respond to a CMS-2567?
Follow the deadline in your survey cover letter. When CMS proposes sanctions for condition-level noncompliance, the notice must allow at least 10 days to respond (42 CFR 493.1810(a)(6)).
What should a CLIA plan of correction include?
For each deficiency: the correction, how you checked scope, root cause, systemic change, monitoring, the responsible person, completion dates and evidence. It must be acceptable to CMS in content and time frames (42 CFR 493.1816).
What happens if a plan of correction is not acceptable?
CMS may cancel the lab's approval to receive Medicare payment if it fails to submit a plan of correction satisfactory to CMS (42 CFR 493.1842).
Which CLIA requirements are conditions?
Requirements identified as conditions in 42 CFR 493.41 and subparts G through Q, such as Successful participation in PT (493.803), Analytic systems (493.1250) and Laboratory director (493.1403, 493.1441).
Can a standard-level deficiency lead to sanctions?
If standard-level deficiencies are not corrected within 12 months after the last day of inspection, CMS cancels Medicare approval and notifies the lab of intent to suspend, limit or revoke its certificate (493.1816(b)).
What is immediate jeopardy in CLIA?
A situation where immediate corrective action is necessary because condition-level noncompliance has caused, is causing, or is likely to cause serious injury, harm or death (42 CFR 493.2).
How fast can CMS suspend a certificate for immediate jeopardy?
If a revisit shows the jeopardy was not eliminated, CMS suspends or limits the certificate no earlier than 5 days after notice (42 CFR 493.1812(b)).
What are the CLIA principal sanctions?
Suspension, limitation or revocation of the CLIA certificate (42 CFR 493.1806).
What are CLIA alternative sanctions?
Directed plan of correction, state onsite monitoring and civil money penalties (42 CFR 493.1806), plus Medicare payment suspension for Medicare-approved labs (493.1807).
Can a waived lab get a civil money penalty?
Alternative sanctions, including civil money penalties, may not be imposed on labs with a Certificate of Waiver (42 CFR 493.1806).
What is unsuccessful proficiency testing?
Under 42 CFR 493.2 it includes unsatisfactory performance for the same analyte in two consecutive or two out of three testing events, among other criteria.
What D-tag is used for unsuccessful PT?
Appendix C directs surveyors to cite unsuccessful PT as a condition-level deficiency using D2016.
What are common CLIA deficiencies?
Frequently reviewed areas include quality control (493.1256), calibration (493.1255), performance specifications (493.1253), procedure manual (493.1251), competency assessment (493.1235), personnel qualifications, proficiency testing and record retention.
How long do I have to appeal a CLIA sanction?
A hearing request must be filed in writing within 60 days from receipt of the notice, unless extended (42 CFR 498.40).
Can I appeal a standard-level CLIA citation?
A finding of standard-level deficiencies in a lab in compliance with condition-level requirements is not an initial determination and is not appealable under 42 CFR 493.1844(c).
Have a citation you need to understand?
Bring the finding. We will help you work out what was cited, what evidence the surveyor expected, and how to fix the system behind it.
Book a free consultation