Book a free consultationTwo terms that sound alike
Per 42 CFR 493.2, unsatisfactory PT performance is failure to attain the minimum satisfactory score for an analyte, test, subspecialty or specialty for a testing event. Unsuccessful participation includes unsatisfactory performance for the same analyte in two consecutive or two out of three testing events, or repeated unsatisfactory overall testing event scores for two consecutive or two out of three events for the same specialty or subspecialty, among other criteria.
Scoring criteria vary by specialty. For example, for general immunology, failure to attain at least 80 percent of acceptable responses for an analyte in a testing event is unsatisfactory analyte performance, and failure to participate in a testing event results in a score of 0 (493.837).
How it is cited
Appendix C tells surveyors that once PT reports confirm unsuccessful PT, "cite as a Condition-level deficiency (use D2016 on the Form CMS-2567)", and that the CMS PT monitoring system may not be used alone to determine unsuccessful participation (SOM Appendix C, §493.803).
First occurrence: training and technical assistance
For an initial unsuccessful performance, CMS may direct training or technical assistance rather than alternative or principal sanctions, unless there is immediate jeopardy, the lab fails to show it has taken steps to correct the problem, or the lab has a poor compliance history (493.803(c); 493.1838). Appendix C says the lab must submit an allegation of compliance listing completion dates for that training or assistance.
Related tags
D2000 (§493.801 enrollment and testing of samples), D2016 (§493.803 successful participation), D5211 (§493.1236 evaluation of PT performance). See the D-tag table.
Frequently asked questions
What is unsuccessful proficiency testing?
Under 42 CFR 493.2 it includes unsatisfactory performance for the same analyte in two consecutive or two out of three testing events, among other criteria.
What D-tag is used for unsuccessful PT?
Appendix C directs surveyors to cite unsuccessful PT as a condition-level deficiency using D2016.
Have a citation you need to understand?
Bring the finding. We will help you work out what was cited, what evidence the surveyor expected, and how to fix the system behind it.